The internal diameter is 1230 mm and length is 1300
mm
The copper table for DUT (Device Under Test) is 700
mm wide and 1200 mm long with M5 holes at 76mm C-C
The following vacuum feed troughs (f.t.) are
available on the movable table for simultaneous use by
DUT:
3 x NW 32
1 x ISO100
1 x ISO160 (on tank)
1 x ISO250
Adapters are available for CF40, CF63 and
DN10-KF
The following vacuum connector flanges with
connectors on both sides are available for use by DUT:
ISO100 with 20 x SMA female-female f.t.
ISO100 with 3 ext. MS3449H22-55P f.t. connected to
3 int. MS3470W22-55S
ISO160 with 50 x SMA female-female f.t.
ISO160 with 1 ext. MS3449H22C-55P f.t. connected
to 1 int. DD50P plus 2 x Lemo SWH.00 f.t.
ISO160 with 1 ext. MS3449H18-32P f.t. connected to
1 int. DC37S and 4 x Lemo ERA1S408CLL E46 (8 kVDC)
f.t.
ISO250 with 10 ext. MS3449H22C-55P f.t. connected
to 10 int. MS3470W22-55S plus 10 x SMA female-female
f.t.
ISO250 with 2 ext. MS3449H22C-55P f.t. and 4 x MHV
(HV BNC) f.t.
MS3449 connectors comply with MIL-C-26482 ser. 2
Pumping time is normally <1 hour to
10-5 mbar at +20° C but depends on
outgassing from DUT
Temperature gradient is ±1° C/min
Pressure is <10-6 mbar at +90°
C
Oil temperature control system for table and shrouds
and N2 system for shrouds
The oil temperature control can independently set
temperatures in the range -35° C to +90° C of
table and shrouds, either under manual or PLC
control
Shrouds can be cooled to -160° C but without
circulation of oil in table, i.e. no control of table
temperature
Additional programmable temperature regulation system
is available for DUT
Solar constant system composed of 4 quartz lamps with
spectral properties simulating the sun. Variable
irradiance from 0-1350 W/m2 within 400 x 400
mm area and 0-1310 W/m2 within 500 x 500
mm
System and DUT temperatures are measured with Pt100
sensors
Vacuum performance better than 10-6mbar at
+25° C and better than 10-5 at +90°
C
PC/Windows based data logging system with 24
channels. Twelve channels monitor the chamber
temperatures, one the chamber vacuum and eleven are
available for monitoring the DUT
Vacuum chamber can be backfilled with either filtered
air or N2
N2 plant with 10 l/h production rate and
3.5 m3 storage tank